• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 12/14 36 +1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 12/14 36 +1.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-51-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision completed due to painful hip.Metal on metal hip revised to ceramic on polyethylene.No surgical delay.Doi: unknown.Dor: (b)(6) 2022.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot: a manufacturing record evaluation (mre) was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 36 +1.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds IN LS11 8DT
UK   LS11 8DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14079621
Report Number1818910-2022-06540
Device Sequence Number17533815
Product Code KWA
UDI-Device Identifier10603295033929
UDI-Public10603295033929
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/11/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1365-51-100
Device Catalogue Number136551100
Device Lot Number2146131
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer04/26/2022
05/05/2022
05/17/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date04/27/2022
05/16/2022
05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT36IDX54OD
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-