ZIMMER BIOMET, INC. DRILL BIT 2.2MM NS; BIT, DRILL Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during an open reduction and internal fixation of a left distal radius fracture, the screwdriver tip broke off inside the proximal locking screw.No intervention has been reported.The fractured piece has been retained in the patient body.No adverse event has been reported as a result of the malfunction.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Reported event was confirmed by review of pictures.Visual examination of the provided pictures identified that the tip of the screwdriver broke off.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT 2.2MM NS
Common Device Name
BIT, DRILL
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16665551
Report Number0001825034-2023-00720
Device Sequence Number17660256
Product Code HWR
UDI-Device Identifier00880304533516
UDI-Public(01)00880304533516(11)211111(10)CA5L246
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number231200200
Device Lot NumberCA5L246
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2023
Supplement Date Received by Manufacturer05/04/2023
Initial Report FDA Received Date04/03/2023
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age21 YR
Patient SexMale
Patient Weight72 KG
-
-