• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ALP2 ALP IFCC GEN.2; NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ALP2 ALP IFCC GEN.2; NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES Back to Search Results
Model Number C503
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable alp2 alkaline phosphatase acc.To ifcc gen.2 results for 1 patient sample on a cobas pro c 503 analytical unit.The initial alp2 result was 337 u/l.The repeat result was 182 u/l.The sample was repeated again on another analyzer and the result was 190 u/l.The initial result was not reported outside of the laboratory.The reagent lot number is 654864 and the expiration date is 31-may-2023.
 
Additional Manufacturer Narrative
The alarm trace did not contain a conspicuous event.Based on the calibration and qc data, we could rule out a general reagent issue.The investigation is ongoing.
 
Additional Manufacturer Narrative
The issue was resolved by changing the reagent and recalibrating the assay.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALP2 ALP IFCC GEN.2
Common Device Name
NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15770603
Report Number1823260-2022-03575
Device Sequence Number17764816
Product Code CJE
UDI-Device Identifier04015630919512
UDI-Public04015630919512
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K033185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberC503
Device Catalogue Number03333752190
Device Lot Number654864
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer11/18/2022
Initial Report FDA Received Date11/10/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-