• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in staten island, new york.Through follow-up communication with the customer, livanova learned that there was neither over-occlusion nor foreign object in the pump raceway.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue, after running for a while with tubing in raceway and water in tubing.In order to solve the issue, motor control (hms) board was replaced and motor power amplifier (hmf) board as a precaution measure.This report was due on november 22, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 19, 2023, before this report was ready to submit.The report was prepared and submitted following restoration of the livanova systems.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova deutschland received a report that a s5 roller pump gave a motor control failure error message during procedure.After turning the pump off and on again the error message disappeared there was no patient injury.
 
Additional Manufacturer Narrative
H10: complaint database analysis revealed that the same pump has been claimed three months earlier for the same error code (mfr reference #(b)(4)).In that occasion, the motor control board (hms) was replaced as precaution, since the root cause was traced back to a pump occlusion being set too high.Based on the information collected, it cannot be ruled out that the re-occurrence of this claimed error code was still related to occlusion setting.The mentioned conclusion is in accordance with the fact that the livanova field service representative did not find any deviation with the pump during testing on site.
 
Event or Problem Description
See initial report.
 
Additional Manufacturer Narrative
Udi related data quality updates only.D2a.Common device name has been corrected in the dedicated section.D2b.Device product code has been corrected in the dedicated section.".
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18299155
Report Number9611109-2023-00626
Device Sequence Number17942410
Product Code DTQ
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)180323
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2023
Supplement Date Received by Manufacturer02/29/2024
08/08/2024
Initial Report FDA Received Date12/11/2023
Supplement Report FDA Received Date03/25/2024
08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-