LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-88-00
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
Livanova received report of a motor control failure error message a pump used on a s5 system during priming.There was no patient involvement.
 
Additional Manufacturer Narrative
Patient identifier-ethnicity: there was no patient involvement.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A faulty pump head cannot be excluded.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Through follow-up communication livanova learned that the affected pump was a s5 mast roller pump.Information related to product information have been updated in dedicated d.1, d.2 , d.4 sections as well as manufacturing date of device in section h.4.H.10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a the motor control board which was replaced to solve the problem.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The root cause of the reported event was a defective motor control board.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key14632145
Report Number9611109-2022-00295
Device Sequence Number18191014
Product Code DWB
UDI-Device Identifier04033817900474
UDI-Public(01)04033817900474(11)200221
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10-88-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date06/08/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-