• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-5537
Medical Device Problem Codes Loss of Osseointegration (2408); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Inadequate Osseointegration (2646)
Date of Event 09/27/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2008 and was revised on (b)(6) 2011.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2008 and was revised on (b)(6) 2011.It is further alleged that he suffered injuries as a result of implantation and explantation of the device at issue.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving an accolade stem was reported.The event was confirmed via medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant indicated: "conclusion/assessment: a patient underwent a tha in 2008 and a revision surgery in 2011.The femoral component was found to be loose at the revision.A 40mm cocr head was used in both cases.Event confirmation: a primary tha with an lfit head can be confirmed.A revision surgery for femoral loosening and use of a cocr head can also be confirmed.Injury and device recall cannot be confirmed.Root cause: the root cause of the primary hip was djd.The root cause for the revision was femoral loosening.Again, injury and/or device recall were not confirmed thus a root cause cannot be ascertained.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to stem loosening.Conclusion/assessment: a patient underwent a tha in 2008 and a revision surgery in 2011.The femoral component was found to be loose at the revision.A 40mm cocr head was used in both cases.Event confirmation: a primary tha with an lfit head can be confirmed.A revision surgery for femoral loosening and use of a cocr head can also be confirmed.Injury and device recall cannot be confirmed.Root cause: the root cause of the primary hip was djd.The root cause for the revision was femoral loosening.Again, injury and/or device recall were not confirmed thus a root cause cannot be ascertained.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 5.5 ACCOLADE (127 DEG) SIZE 5.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7911690
Report Number0002249697-2018-03073
Device Sequence Number19591830
Product Code LPH
UDI-Device Identifier04546540510778
UDI-Public04546540510778
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2011
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/20/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2011
Device Catalogue Number6021-5537
Device Lot Number16788403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2018
Supplement Date Received by Manufacturer07/25/2024
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/15/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
-
-