• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AURORA SPINE, INC. PEEK SPACER INTRODUCER HANDLE; SPINOUS PROCESS PLATE, SPINAL INTERLAMINAL FIXATION ORTHOSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AURORA SPINE, INC. PEEK SPACER INTRODUCER HANDLE; SPINOUS PROCESS PLATE, SPINAL INTERLAMINAL FIXATION ORTHOSIS Back to Search Results
Catalog Number 108-02610-0810
Device Problems Break (1069); Difficult to Remove (1528); Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/21/2015
Event Type  malfunction  
Event Description
During transforaminal lumbar interbody fusion at l4-l5, the surgeon attempted to place an aurora titanium coated peek spacer, however the insertion handle was unable to be removed from the peek spacer.Decision was made at that point to transition to a different interbody spacer.Introducer on engagement to the implant, locked and would not release.After manipulation to remove, it broke in the hands of the surgeon.All pieces removed from surgical field, no harm to patient, no retained pieces as verified by x-ray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEK SPACER INTRODUCER HANDLE
Type of Device
SPINOUS PROCESS PLATE, SPINAL INTERLAMINAL FIXATION ORTHOSIS
Manufacturer (Section D)
AURORA SPINE, INC.
1920 palomar point way
carlsbad CA 92008
MDR Report Key4486492
MDR Text Key5497909
Report Number4486492
Device Sequence Number2
Product Code PEK
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial
Report Date 01/29/2015
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Invalid Data
Device Catalogue Number108-02610-0810
Device Lot Number125-22
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/29/2015
Event Location Hospital
Date Report to Manufacturer02/05/2015
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/29/2015
Patient Sequence Number1
Patient Age29 YR
Patient Weight74
-
-