SORIN GROUP USA HEART-LUNG MACHINE Back to Search Results
Model Number S5
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Abnormal Blood Gases (1034)
Event Date 05/09/2017
Event Type  Injury  
Event Description
High pco2 levels noted during cardiopulmonary bypass by poc blood gas analyzer.Values were rechecked with the same result.Perfusionist started troubleshooting the equipment.The vaporizer was replaced and values returned to normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART-LUNG MACHINE
Type of Device
HEART-LUNG MACHINE
Manufacturer (Section D)
SORIN GROUP USA
arvada CO
MDR Report Key6636313
MDR Text Key77524536
Report NumberMW5070353
Device Sequence Number2
Product Code DTQ
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/24/2017
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS5
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/09/2017
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-