At the end of the purging air sequence during an extracorporeal photopheresis (ecp) treatment, the bowl dislodged from the centrifuge platen and shattered on impact with the centrifuge chamber walls.The procedure was immediately aborted with 152 ml of blood loss to the patient.Despite the centrifuge chamber door being closed and locked, blood splatter from the bowel shattering reached the ceiling, nearby equipment (computer, iv pump, vs monitor), supply cart, curtains at the far end of the bay and the walls.The instrument was pulled from service and will be cleaned and repaired by the manufacturer's contracted service company.This was the second shattered bowl incident in 3 days, using the same lot number of procedural kit.This lot was removed from inventory, and the manufacturer was notified.
|