• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 11MM/130 DEG TI CANN TFNA 235MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 11MM/130 DEG TI CANN TFNA 235MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.037.145S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.This report is for an unknown femoral nail/unknown lot.Part and lot number are unknown.Without the specific part number; the udi number and 510-k number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from depuy synthes reports an event in germany as follows: it was reported that on an unknown postoperative date a femoral nail fractured.No further information provided.This report is for one (1) unknown femoral nail this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the tfna fem nail ø11 le 130° l235 timo15 was cracked from one of the walls of the blade insertion hole and broken on the other side.Breakage of the nail cannot be confirmed since the fragment is still attached to the nail.There are no signs of drilling along the section of the blade insertion hole, meaning that the it was implanted according to the surgical technique guide and the edges of the fragments do not indicate any material issues, hence these are not considered to be potential factors for the identified defect.X-ray images from (b)(6) 2022 show a nail and blade construct at the left proximal femur with signs of an intertrochanteric fracture.Revision x-ray from (b)(6) 2022 shows the nail being fractured and the bone not completely healed.The patient presents extensive vascular calcifications which may indicate an underlying disease, process that could negatively influence the potential for bone union, this observation and a possible early weight-bearing could have led to failure of the implant.The tfn-advanced¿ proximal femoral nailing system (tfna) surgical technique guide was reviewed.Following relevant statements were found.Precautions: compromised vascularity in the site of proposed implantation may prevent adequate healing and thus preclude the use of this or any orthopaedic implant.A dimensional inspection for the tfna fem nail ø11 le 130° l235 timo15 was not performed due to post manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the tfna fem nail ø11 le 130° l235 timo15 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Manufacturing location: monument manufacturing date: 02-jun-2022, expiration date: 01-may-2032, part number: 04.037.145s, 11mm/130 deg ti cann tfna 235mm/left-sterile, lot number: 851p324 (sterile), lot quantity: (b)(4).This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: part number: 04.037.912.3, tfna lock drive, lot number: 827p844, lot quantity: unknown.A dhr could not be accessed for this review.Part number: 04.037.942.2, lock prong, 130 degree tfna, lot number: 851p602, lot quantity: (b)(4).Production order traveler met all inspection acceptance criteria.Part number: 21127, timoagri16.00, lot number: 538p353, lot quantity: (b)(4).Inspection instruction met all inspection acceptance criteria.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from the photo.Visual analysis of the photo revealed that there was no damage or defects, the photos do not provide enough evidence to confirm the reported condition.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the shaft was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11MM/130 DEG TI CANN TFNA 235MM/LEFT - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key15792783
Report Number8030965-2022-09790
Device Sequence Number2009656
Product Code HSB
UDI-Device Identifier10886982096579
UDI-Public(01)10886982096579
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.037.145S
Device Catalogue Number04.037.145S
Device Lot Number851P324
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer12/14/2022
01/31/2023
Initial Report FDA Received Date11/14/2022
Supplement Report FDA Received Date01/06/2023
02/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-