SENSEONICS INC. EVERSENSE TRANSMITTER; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102208-500
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, senseonics was made aware of an incident where user experienced false hyperglycemia due to inaccuracies in sensor readings.
 
Additional Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Additional Manufacturer Narrative
The complaint was investigated, and though the reported sg and bg values did not match with that in the data management system (dms) except for one bg value, mismatches between the bg and sg values in the early days of the sensor might be due to the initial variability of the system, as the calibrations were still helping the system adjust.Per the case notes of associated sensor inaccuracies complaint (sf #00076383), there is a better agreement between sg and bg values and user is using the system with up to date information.No further investigation was found necessary for this complaint.User wasn't symptomatic, didn't seek medical treatment and didn't need to self treat as no symptoms were shown.As user's hcp was not aware of the event, user was advised to let hcp know.Per dms, user is currently using the system with up to date information.D2.Product code updated to qhj.H3.Device evaluated by manufacturer? yes.H6.Investigation findings changed to 114.H6.Investigation conclusion changed to 4315.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE TRANSMITTER
Common Device Name
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key13932552
Report Number3009862700-2022-00048
Device Sequence Number20292821
Product Code QHJ
UDI-Device Identifier00817491021441
UDI-Public00817491021441
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P160048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/05/2022
Device Model Number102208-500
Device Catalogue NumberFG-3300-01-001
Device Lot Number123461
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-