WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 1555200500
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal fusion t herapy.It was reported that the rod was broken from the initial surgery and was replaced it with a new rod. there were no further complications  or symptoms reported regarding the event.Additional information was received via manufacturer representative that there were no broken fragments left inside the patient.
 
Additional Manufacturer Narrative
H3: product analysis 0f product#1555200500, lot# 0793172w visual inspection confirmed the rod has been broken.Severe damage and smearing noted on fracture surfaces.No pre-existing surface defects were identified that could contribute to the breaks.After visual, microscopic, no evidence was found that would suggest a defect in manufacturing or processing of the implants.This type of damage is consistent with failure due to bend stress overload.Radiographic image review: single ap xray thoracic, pedicle screw and rods are present.The rod appears to be fractured on one side.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key17236644
Report Number1030489-2023-00444
Device Sequence Number20305254
Product Code NKB
UDI-Device Identifier00643169272255
UDI-Public00643169272255
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131321
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1555200500
Device Catalogue Number1555200500
Device Lot Number0793172W
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date09/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age17 YR
Patient SexFemale
-
-