• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 19MM CALCAR BODY/BOLT (+30) COMPONENT LEVEL 9002-0-5731; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 19MM CALCAR BODY/BOLT (+30) COMPONENT LEVEL 9002-0-5731; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6276-4-319
Medical Device Problem Codes Degraded (1153); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Injury (2348); Reaction (2414); Metal Related Pathology (4530)
Date of Event 07/19/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Device evaluated by manufacturer? not returned.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with and lfit anatomic cocr v40 femoral head on her right hip on or about (b)(6) 2010 and was revised on (b)(6) 2017.It is further alleged that she suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with and lfit anatomic cocr v40 femoral head on her right hip on or about (b)(6) 2010 and was revised on (b)(6) 2017.It is further alleged that she suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level involving a restoration modular body was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event could not be confirmed as insufficient information was provided.Further information such as device return, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
19MM CALCAR BODY/BOLT (+30) COMPONENT LEVEL 9002-0-5731
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key8753000
Report Number0002249697-2019-02486
Device Sequence Number20359947
Product Code LPH
UDI-Device Identifier07613153071594
UDI-Public07613153071594
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K050138
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/17/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2012
Device Model Number6276-4-319
Device Catalogue Number6276-4-319
Device Lot NumberT55352BC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2019
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date07/02/2019
Supplement Report FDA Received Date04/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-