• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. FMC CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. FMC CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number 050-50758
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
A peritoneal dialysis (pd) patient contacted fresenius technical support to report the liberty select cycler alarmed with an m31 air detected in cassette warning during drain 1 of 5.A fluid leak was subsequently discovered in step 9 of treatment complete.The patient was not connected during the incident.Fluid was found in both the pump module area and on the external portion of the cassette.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon followup the clinic manager (cm) at the patient¿s clinic confirmed the reported event.The cm stated that patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cm could not confirm if the patient completed treatment on the day of the reported event.The cm confirmed the patient has continued using the cycler for their pd treatment without issue since.The cm confirmed that the cycler set was discarded and not available to be returned to the manufacturer for physical evaluation.No defect or damage to the cycler set cassette was noted upon removing it from the cycler.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Additional Manufacturer Narrative
Correction: d1, d4, g4 pma/510(k)#, h4 plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
A peritoneal dialysis (pd) patient contacted fresenius technical support to report the liberty select cycler alarmed with an m31 air detected in cassette warning during drain 1 of 5.A fluid leak was subsequently discovered in step 9 of treatment complete.The patient was not connected during the incident.Fluid was found in both the pump module area and on the external portion of the cassette.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.Upon followup the clinic manager (cm) at the patient¿s clinic confirmed the reported event.The cm stated that patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cm could not confirm if the patient completed treatment on the day of the reported event.The cm confirmed the patient has continued using the cycler for their pd treatment without issue since.The cm confirmed that the cycler set was discarded and not available to be returned to the manufacturer for physical evaluation.No defect or damage to the cycler set cassette was noted upon removing it from the cycler.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMC CASSETTE
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key17243830
Report Number0008030665-2023-00588
Device Sequence Number20431082
Product Code FKX
UDI-Device Identifier00840861100699
UDI-Public00840861100699
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/31/2025
Device Model Number050-50758
Device Catalogue NumberUNKNOWN- FMC CASSETTE
Device Lot Number22SR08083
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2023
Supplement Date Received by Manufacturer08/07/2023
Initial Report FDA Received Date06/30/2023
Supplement Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; DELFLEX PD FLUID; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER
Patient Age61 YR
Patient SexMale
Patient Weight72 KG
-
-