Model Number CTS02
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Name of procedure being performed: ni.Detailed description of event: when the item was being repackaged at the start of surgery, it was noticed that there was a hair inside the pouch.Due to it being a biohazard, the tech placed the trocar back in the packaging and another, new trocar was used to complete the case.Product is available for return.Patient status: na (no patient involvement).Type of intervention: the device was placed back in package and a new one was used to complete the case.
 
Additional Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.Correction: d2 - device product code was corrected to gcj as this was initially provided incorrectly.
 
Event or Problem Description
Name of procedure being performed: ni.Detailed description of event when the item was being unpackaged at the start of surgery, it was noticed that there was a hair inside the pouch.Due to it being a biohazard, the tech placed the trocar back in the packaging and another, new trocar was used to complete the case.Product is available for return.Patient status: na (no patient involvement).Type of intervention: the device was placed back in package and a new one was used to complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTS02, 5X100 KII SLEEVE ZTHR 12/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15953903
Report Number2027111-2022-00854
Device Sequence Number20531194
Product Code GCJ
UDI-Device Identifier00607915123499
UDI-Public(01)00607915123499(17)250830(30)01(10)1462084
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K060096
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/30/2025
Device Model NumberCTS02
Device Catalogue Number101208701
Device Lot Number1462084
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date12/09/2022
Supplement Report FDA Received Date02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1