STRYKER ORTHOPAEDICS-MAHWAH MRH KNEE FEM S RGT; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 64811111
Medical Device Problem Codes Fracture (1260); Loss of Osseointegration (2408)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/31/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's right hinged knee was revised after patient complaint of thigh pain.Surgeon originally suspected femoral loosening.However, intra-op it was observed that the lateral condyle of the femoral component broke off of the implant and was loose in the joint.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture and loosening involving a mrh femoral component was reported.The event was confirmed for crack/fracture.Method & results: product evaluation and results: the reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted femoral component with fractured condyle.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient had pain; intraop it was observed that the implant was fractured and loose.The reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted femoral component with fractured condyle.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's right hinged knee was revised after patient complaint of thigh pain.Surgeon originally suspected femoral loosening.However, intra-op it was observed that the lateral condyle of the femoral component broke off of the implant and was loose in the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRH KNEE FEM S RGT
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17805312
Report Number0002249697-2023-01071
Device Sequence Number20571378
Product Code KRO
UDI-Device Identifier07613327045147
UDI-Public07613327045147
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K994207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/18/2023
Device Catalogue Number64811111
Device Lot NumberDP63C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/31/2023
Supplement Date Received by Manufacturer11/10/2023
Initial Report FDA Received Date09/24/2023
Supplement Report FDA Received Date12/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-