• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS CA 19-9; TEST, CARBOHYDRATE ANTIGEN (CA19-9)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS CA 19-9; TEST, CARBOHYDRATE ANTIGEN (CA19-9) Back to Search Results
Model Number CA 19-9
Medical Device Problem Codes Incorrect Measurement (1383); High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys ca 19-9 immunoassay results for 1 patient sample on a cobas e 411 immunoassay analyzer.The initial ca 19-9 result was > 1000 u/ml.A 1:5 dilution was made with the patient sample and the result was 242.2 u/ml.The sample was run undiluted again and the result was > 1000 u/ml.A 1:2 dilution was then made with the patient sample and the result was 481.4 u/ml.The sample was recentrifuged and run undiluted again and the result was > 1000 u/ml.No questionable results were reported outside of the laboratory.The reagent lot number is 598103.The expiration date was requested but not provided.
 
Additional Manufacturer Narrative
The patient is diagnosed with asthma.The investigation confirmed the customer's dilution results.Three-hour neuraminidase treatment was performed on the patient sample and it showed incomplete degradation.Protein a addition treatment was performed and the result showed 1.5% recovery.Neuraminidase treatment was also performed on a competitor analyzer and the result showed complete degradation.The investigation determined that the root cause of the discrepant dilution results is consistent with sample interference.Product labeling states, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design." the investigation did not identify a product problem.Based on the available data, a general reagent issue could be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS CA 19-9
Common Device Name
TEST, CARBOHYDRATE ANTIGEN (CA19-9)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15836483
Report Number1823260-2022-03713
Device Sequence Number20689517
Product Code NIG
UDI-Device Identifier04015630904297
UDI-Public04015630904297
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050231
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCA 19-9
Device Catalogue Number11776193122
Device Lot Number598103
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-