• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN TRIGEN SURESHOT DISTAL TARGETING SYST DRILL BIT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN TRIGEN SURESHOT DISTAL TARGETING SYST DRILL BIT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 71440209
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a tka surgery, the gii slt 3 mod tib cutblck lt while drilling holes, the drill bit broke off inside the hold.Surgery was resumed, without any delay, with a smith and nephew back-up device.Patient was not harmed.No further complications were reported.
 
Additional Manufacturer Narrative
Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be broken and lodged within the cutting block, rendering the device inoperative.The rest of the drill bit was not returned.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A contribution of the device to the reported event could be corroborated as the device shows signs of damage/wear.Possible causes could include but not limited to surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that, during a tka surgery, the gii slt 3 mod tib cutblck lt while drilling holes, the trigen sureshot distal targeting syst drill bit broke off inside the hold.Surgery was resumed, without any delay, with a smith and nephew back-up device.Patient was not harmed.No further complications were reported.
 
Additional Manufacturer Narrative
Corrected data: h6 - code 2199 removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN TRIGEN SURESHOT DISTAL TARGETING SYST DRILL BIT
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13926977
Report Number1020279-2022-01357
Device Sequence Number2148548
Product Code OLO
UDI-Device Identifier03596010460431
UDI-Public03596010460431
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/06/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71440209
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer04/08/2022
04/20/2022
04/20/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/12/2022
04/21/2022
12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/17/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71440209/15GM11047 - GII SLT 3 MOD TIB CUTBLCK LT
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-