• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 9/10 COCR HEAD 28MM +6; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 9/10 COCR HEAD 28MM +6; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 152190060
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.B3: date of event is an unknown date.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent the primary surgery via tha with the liner in question.On an unknown date, it was confirmed the liner in question was worn out.Therefore, revision surgery will be scheduled on (b)(6) 2023.Only the liner will be replaced, but the cup and stem will also be replaced depending on the status of the other implants.If only the liner and head are replaced, use jj product implants; if the replacement extends to the cup and stem, use products made by another company (zb).It is unknown whether the implants other than reported products used in the primary surgery are jj products or not.No further information is available.New information was received from sales rep on dec 21,2023 as follows: the primary surgery was performed on (b)(6) 2002, at another hospital.There was wear of the polyethylene over time, but the final reason for the revision surgery was head-neck junction failure since the metal head was de-rotated from the aml plus stem.This is all that is known about the current situation before revision surgery is performed.New information received from the sales rep on dec 28, 2023 as follows: revision surgery was performed on (b)(6) 2023 at (b)(6) hospital and all implants were removed.After removal of the stem and liner, the screw and cup were removed after checking the fixation of the cup and to ensure compatibility.The revision procedure was then completed with a zb cementless cup and an oss (orthopedic salvage system) stem.A simple check of the stem and head after removal showed that the fixation function of the neck junction had failed.This was probably due to metal wear of the neck junction.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Although the cup was not defective, after removal of the stem and liner, the screw and cup were removed after checking the fixation of the cup and to ensure compatibility.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM +6
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18415475
Report Number1818910-2023-26125
Device Sequence Number2761085
Product Code JDI
UDI-Device Identifier10603295276654
UDI-Public10603295276654
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/28/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2006
Device Catalogue Number152190060
Device Lot NumberXYW-90
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer12/28/2023
01/10/2024
Initial Report FDA Received Date12/28/2023
Supplement Report FDA Received Date01/03/2024
01/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/23/2001
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDURON 10D 56 OR 68ODX28ID.; UNKNOWN HIP FEMORAL STEM.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-