CARDINAL HEALTH 200, LLC ARGYLE; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
|
Back to Search Results |
|
Model Number 8888505040 |
Device Problem
Scratched Material (3020)
|
Patient Problem
Ulcer (2274)
|
Event Type
No Answer Provided
|
Event Description
|
I am an anesthesiologist and have had 3 patients end up with wounds in the back of their mouths after using this suction tip.The first 2 patients i did not realize how dangerous it was.The 3rd patient, i very gingerly used the suction tip prior to extubating the patient.The suction tip has holes on the sides near the tip that are cut so that the edges of the holes are sharp.You can feel the sharp edges with your finger and fingernail.I have shown the suction tip to an ent surgeon i work with and he replied "this should not to placed anywhere in the human body".I have attached pictures of the packaging material as well as a closeup of the tip of the suction tip.
|
|