| Catalog Number |
4566 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Inflammation (1932)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented an inflammation 6 weeks after a surgery where 2 regeneten devices were used.The patient was brought back to surgery at 8 weeks, no infection was found but the doctor felt the regeneten was ¿mushy¿ and decided to remove it.He was not sure if there was an issue with regeneten, but he is also not sure it was incorporating like it should.Fixation could have also been an issue as the distal part of the implant may not have been fixated well.No additional complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h2: additional information on b5.H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented an inflammation 6 weeks after a surgery where a regeneten implant was used.The patient was brought back to surgery at 8 weeks, no infection was found but the doctor felt the regeneten was ¿mushy¿ and decided to remove it.He was not sure if there was an issue with regeneten, but he is also not sure it was incorporating like it should.Fixation could have also been an issue as the distal part of the implant may not have been fixated well.No additional complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction in d2 and g4 (510k).
|
| |
|