DEPUY IRELAND - 9616671 PFC* FLUT TIB ROD TRL 75 X 14; KNEE INSTRUMENT : TIBIAL STEM TRIALS Back to Search Results
Model Number 96-6876
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
During a knee revision a zimmer lcck was taken out and a tc3 was going to be put in.During the preparation of tibia they had reamed with the 14 mm reamer to the right depth and with the conical reamer and went on to broach with the smallest sleeve broach and the 14 x 75 mm stem attached.During the first broach the tip of the trial stem came off the trial stem and got stuck distally in the patients tibia.It was no chance of retrieving it because it was stuck.The tip of the stem is not possible to loosen but it came off anyway.There was a surgical delay of 45 minutes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : visual examination of the returned device confirmed the reported event.The device was found broken, not all parts were returned.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mrae) was not performed.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE INSTRUMENT : TIBIAL STEM TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14468721
Report Number1818910-2022-09366
Device Sequence Number4570161
Product Code HWT
UDI-Device Identifier10603295217589
UDI-Public10603295217589
Combination Product (Y/N)N
Initial Reporter CountrySW
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-6876
Device Catalogue Number866876
Device Lot NumberTBFCV
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/14/2022
07/19/2022
Initial Report FDA Received Date05/23/2022
Supplement Report FDA Received Date06/16/2022
07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2001
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
Patient Weight80 KG
-
-