• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR PLUS; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR PLUS; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 261P
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2019
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Additional Manufacturer Narrative
The manufacturer previously reported life-threatening, hospitalization - initial or prolonged in b.2 section, which was marked incorrectly as this is product problem only as mentioned in b.1 section in previous report.G.6 section incorrectly marked 30 days and initial including 15 days, which should marked as 30 days and initial only in this follow up report.E.1 section was missing in previous report, which is correctly updated in this follow up report.Common device name in d.2 section was incorrectly reported as mns in previous report, which is correctly reported now in this follow up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR PLUS
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13660692
Report Number2518422-2022-07884
Device Sequence Number5009485
Product Code BZD
UDI-Device Identifier00606959028654
UDI-Public00606959028654
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/17/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number261P
Device Catalogue Number261P
Was Device Available for Evaluation? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 09/26/2019
Supplement Date Received by Manufacturer09/26/2019
Initial Report FDA Received Date03/03/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Disability; Hospitalization;
Patient SexUnknown
-
-