• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 12/14 40 +8.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 12/14 40 +8.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-07-100
Medical Device Problem Codes Degraded (1153); Naturally Worn (2988)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Adhesion(s) (1695); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Synovitis (2094); Anxiety (2328); Distress (2329); Discomfort (2330); Depression (2361); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Fibrosis (3167); Metal Related Pathology (4530); Unspecified Tissue Injury (4559)
Date of Event 12/03/2021
Type of Reportable Event Serious Injury
Event or Problem Description
(b)(6) litigation record received.Litigation alleges heavy metal poisoning from the toxic heavy metals resulting to muscle, tissue and bone injury, metallosis, metal wear, scar tissue formation, pain and limited adl.Also suffered emotional trauma and distress.Doi: (b)(6) 2010; dor: (b)(6) 2021; right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
After review, patient was revised due to metallosis with synovitis.Evidence of corrosion at trunnion of stem and at liner cup interface.Posterior soft tissue sleeve and abductors intact.Large osteolytic defect in right femur greater trochanter.Components well fixed and in acceptable position.Osteolysis around superior rim of acetabular component.However, there was no gross evidence of infection.There was gross evidence of trunnionosis with black grey corrosion products evident.The acetabular liner was removed revealing corrosion behind the 40 mm cobalt chrome liner.Doi: (b)(6) 2010 dor: (b)(6) 2021 affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary; no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot; a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.H10 additional narrative: added: a2 (dob), a3, b5, h6 (clinical codes) corrected: d6a.
 
Event or Problem Description
Medical records received.In addition to what was previously reported in the medical records the patient was revised to address depression, walking difficulty and metallosis.Clinical visit reported persistent focus of ossification of the right greater trochanter.Surgical pathology reported tissue reaction to particulate implant material and in the soft tissue metallic particulate debris and corrosion products.Alhough there is abundant corrosion product and metallic pigment with synovial tissue, there is not a significant alval indicative of an immunogenic reaction.Femoral head shows circular abrasive wear.Microscopic description demonstrate interface synovial like tissue showing dense and diffuse sub synovial fibrosis and are foci of reactive appearing synovial hyperplasia.Mri hips shows pseudotumor, elevated metal ions, lateral pain and bursitis.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.All available x-rays evidence were reviewed, and nothing indicative of a device nonconformance was observed.Therefore, the reported allegation was not confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed, as no lot number was retrieved for this device.
 
Event or Problem Description
The patient was revised to address osteolysis, stiffness, limited mobility, adl, and anxiety.Head and liner were exchanged and femoral bone allographs were performed.The operative note reported metallosis with synovitis.There was evidence of corrosion at the trunnion stem and liner cup interface.Large osteolytic defect in the right femur greater trochanter.Osteolysis around the acetabular component.There was evidence of trunnions with black grey corrosion in the head.Scar tissues were excised.Clinical visit reported mild tenderness over the greater trochanter, and mild nonspecific synovial expansion more evident on the right side.There are hypointense foci likely related to metal debris and osteolysis in the proximal femur, metallosis, adverse local tissue reaction, and elevated metal ions.However, the lab result is below 7 ppb.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.All available x-rays evidence were reviewed, and nothing indicative of a device nonconformance was observed, therefore the reported allegation was not confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was retrieved for this device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary = no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot =a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.H10 additional narrative: added: h6 (clinical code).H6 clinical codes: metal related pathology (e1618) is used to capture metal poisoning and blood heavy metal increased.Unspecified tissue injury (e2015) is used to capture soft tissue injury and bone injury.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records were reviewed by a clinician and states that the patient underwent a right total knee arthroplasty utilizing depuy synthes implants in (b)(6) 2014.There were no intraoperative or postoperative complications noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 40 +8.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds IN LS11 8DT
UK   LS11 8DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14874969
Report Number1818910-2022-12075
Device Sequence Number5194884
Product Code KWA
UDI-Device Identifier10603295033011
UDI-Public(01)10603295033011(17)140614(10)2940273
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 06/30/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2014
Device Model Number1365-07-100
Device Catalogue Number136507100
Device Lot Number2940273
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/04/2022
03/13/2023
03/28/2023
04/18/2023
05/22/2023
06/12/2023
08/17/2023
10/13/2023
08/08/2024
Initial Report FDA Received Date06/30/2022
Supplement Report FDA Received Date07/06/2022
03/14/2023
04/06/2023
04/18/2023
06/06/2023
06/12/2023
08/25/2023
10/16/2023
08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/19/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; PINNACLE 100 HA ACET CUP 56MM; PINNACLE MTL INS NEUT40IDX56OD; SUMMIT DUOFIX TAP SZ5 HI OFF; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP ACETABULAR LINER METAL PINNACLE; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight94 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-