• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170052
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 12/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, at the final steps of closing the peritoneum, the jaws of the device were closed and the surgeon felt a clunk.When the surgeon withdrew the device, the needle was broken.The broken portion of the needle fell into the patient's cavity and an x-ray was performed, but the remaining portion of the needle was not found.The surgical time was extended due to the product problem.
 
Additional Manufacturer Narrative
Concomitant medical products: 173016 endo stitch instrument (lot#: unknown).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Additional information: h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the needle was broken or damaged.It was reported that the needle broke into separate pieces, a component disengaged from the device into the surgical cavity and operating room time was extended by more than 30 minutes resulting from product failure.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.These issues can occur if the needle was not loaded properly or when the needle was forced into an obstacle.This condition can also occur if excess pressure was exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the toggle levers are fully retracted prior to opening the jaws of the device.The reloading button should never be pressed when the instrument is in the body cavity, as this will release the needle.Inspect the application site to ensure hemostasis.Place additional sutures or use electrocautery if necessary to complete hemostasis.Endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.A thorough understanding of the operating principles, risks versus benefits, and the hazards involved in utilizing an endoscopic approach is necessary to avoid possible injury to the user or patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
According to the reporter, during a laparoscopic sleeve gastrectomy procedure, at the final steps of closing the peritoneum, the jaws of the device were closed and the surgeon felt a clunk.When the surgeon withdrew the device, the needle was broken.The broken portion of the needle fell into the patient's cavity and an x-ray was performed, but the remaining portion of the needle was not found.The surgical time was extended more than 30 minutes due to the product problem.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Additional information: g4: (510k), d4: (model and catalog number, udi).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16229992
Report Number9612501-2023-00141
Device Sequence Number5531139
Product Code OCW
UDI-Device Identifier20884521126746
UDI-Public20884521126746
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/31/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number170052
Device Catalogue Number170052
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer01/24/2023
02/09/2023
03/20/2023
Initial Report FDA Received Date01/23/2023
Supplement Report FDA Received Date02/09/2023
02/24/2023
03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-