| Model Number |
115-222 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Seroma (2069)
|
| Date of Event |
03/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, opening, crease fold, wear abrasion and red particles in the surface of the shell.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a opening sharp in the posterior side assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Patient reported right side rupture.Healthcare professional additionally reported via mri ¿can found small amount of seroma around device.¿ the device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported via mri ¿can found small amount of seroma around device.¿ the device has been explanted.
|
| |
|
Search Alerts/Recalls
|