ENCORE MEDICAL L.P. BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA OPTION HEAD, 40MM Back to Search Results
Model Number 942-01-40G
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 04/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to dislocation.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgerydetailed in this event occurred 42 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions or incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HIP
Common Device Name
HEAD, FEMORAL, CERAMIC, BILOX DELTA OPTION HEAD, 40MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14376051
Report Number1644408-2022-00627
Device Sequence Number5782527
Product Code LZO
UDI-Device Identifier00888912076265
UDI-Public(01)00888912076265
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082844
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number942-01-40G
Device Catalogue Number400-04-400
Device Lot Number884B1133
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date05/12/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-04-400 LOT 884B1133; 400-05-300 LOT 889B1140; 400-05-300 LOT 889B1140; 940-00-030 LOT 147Z1103; 940-00-030 LOT 147Z1103; 940-00-035 LOT 148Z1104; 940-00-035 LOT 148Z1104; 940-02-54G LOT 801Z1150; 940-02-54G LOT 801Z1150; 942-01-40G LOT 628Z1048; 942-01-40G LOT 628Z1048; 942-01-40G LOT 628Z1064
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexUnknown
-
-