| Model Number |
1365-51-100 |
| Medical Device Problem Codes |
Degraded (1153); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Synovitis (2094); Distress (2329); Metal Related Pathology (4530)
|
| Date of Event |
09/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Surgeon stated that patient was presented to clinic with pain.Removed pinnacle metal on metal liner, as well as two screws and cocr femoral head and replaced with poly liner and ceramic head with ts sleeve.Doi: unknown, dor: (b)(6) 2022, affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).The device catalog number is unknown; therefore, udi is unavailable.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Litigation records received.Patient was revised was due to heavy metal poisoning from the toxic heavy metals.Records also alleges injury, pain, metallosis, metal wear, loss of enjoyment of life, limitation of daily activities, emotional trauma and distress.Doi: (b)(6), 2010.Dor: (b)(6) 2022.Right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b1 (product problem), b5, d4, d6a, g2, g4, h4, h6 (clinical codes).H6 clinical code: metal related pathology (e1618) is used to capture metal poisoning and blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2a, d2b, d10, e1, e2, e3, g1, h6 (medical device problem code).E3 initial reporter occupation: lawyer.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : device associated with this report was not received for examination.All available x-rays were reviewed and no defect was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|
Event or Problem Description
|
|
After a review of the medical records, the patient was revised to address the complication of metal on metal right total hip replacement resulting in pain and elevated metal ions.The operative note reported scar tissue, opaque fluid, synovitis, and dark staining consistent with metallosis.There was minimal oxidative debris at the junction of the trunnion stem.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: device associated with this report was not received for examination.All available x-rays were reviewed and no defect was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary; device associated with this report was not received for examination.All available x-rays were reviewed and no defect was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot; the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|