MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number VBJR061502A
Medical Device Problem Codes Failure to Deliver (2338); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Stenosis (2263); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium apl, which in covalently bound to the device surface and is essentially|non-eluting.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2021, a 6mm x 15cm gore® viabahn® endoprosthesis (vsx) device was intended for use in the superficial femoral artery for treatment of stenosis.After the device was advanced across the lesion the physician attempted to deploy the vsx device.However, after the deployment line was removed from the patient, the physician attempted to remove the delivery catheter and felt resistance.An angiograph showed approximately 10 centimeters of the vsx device had full expansion and the remaining device was still constrained.The physician attempted to capture the device in the introducer sheath but was unsuccessful.The physician decided to pull the delivery catheter and vsx device to the access site, and perform a cut down to remove the vsx device.An additional vsx device was successfully delivered to complete the procedure.The patient did not experience any adverse consequences.
 
Additional Manufacturer Narrative
Section h6 updated to reflect completion of investigation.Manufacturing records were reviewed, and the device met all pre-release specifications.The endoprosthesis was returned.The cause of the device failure mode could not be determined.Cause of the reported event cannot be established based on evaluation of the returned delivery system and the information reported to gore.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.
 
Additional Manufacturer Narrative
Revised b1 to capture adverse event and product problem - yes.Revised d2 product code.Revised g4 pma/510(k)number.Revised h6 health effect - clinical code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
nick lafave
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key12847775
Report Number2017233-2021-02552
Device Sequence Number6212738
Product Code NIP
UDI-Device Identifier00733132623747
UDI-Public00733132623747
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/12/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2022
Device Model NumberVBJR061502A
Device Catalogue NumberVBJR061502A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/12/2021
Supplement Date Received by Manufacturer06/01/2022
08/11/2022
Initial Report FDA Received Date11/19/2021
Supplement Report FDA Received Date06/01/2022
08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexMale
-
-