• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 56 MM O.D.; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. SHELL POROUS WITH CLUSTER HOLES 56 MM O.D.; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Malposition of Device (2616); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Implant date: unknown day in (b)(6) 2021.The device will not be returned for analysis, as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2022 - 00097, 0001822565 - 2022 - 01122, 0001822565 - 2022 - 01123.
 
Event or Problem Description
It was reported the patient underwent a right hip procedure.Subsequently, the patient states to have had pain in the groin and both legs since the procedure.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: ref (b)(4); lot #3020958; ceramic femoral head.
 
Event or Problem Description
It was reported the patient underwent a right hip procedure.Subsequently, the patient states to have had pain in the groin and both legs since the procedure.The doctor told the patient that the implants were installed crooked.No revision has been reported to date.No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: relatively normal appearance of the bilateral total hip arthroplasties.It is noted that there is slight lateral eccentric positioning of the femoral tips in both of the arthroplasties, nonspecific and potentially related to initial surgical technique.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The root cause remains the same.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.Part and lot identification are necessary for review of device history records, neither were provided.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHELL POROUS WITH CLUSTER HOLES 56 MM O.D.
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14137832
Report Number0001822565-2022-01121
Device Sequence Number6254184
Product Code LPH
UDI-Device Identifier00889024118157
UDI-Public(01)00889024118157(17)300712(10)64670038
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K934765
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/28/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00620005622
Device Lot Number64670038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer05/03/2022
05/06/2022
06/21/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date05/10/2022
05/24/2022
06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight73 KG
-
-