• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 URINE COLLECTION; DRAIN BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 URINE COLLECTION; DRAIN BAG Back to Search Results
Model Number A902916
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the customer had multiple complaints about the surestep foley tray system that they had in stock.Stated that the main concern was foley bag not draining effectively and the nurses have to put the bag on the floor in order to get it to drain.It was stated that this was a concern when doing surgery as the provider were finding full bladders possible leading to one injury so far.While putting the bag on the floor was helping, it was absolutely against all infection control standards.
 
Event or Problem Description
It was reported that the customer had multiple complaints about the surestep foley tray system that they had in stock.Stated that the main concern was foley bag not draining effectively and the nurses have to put the bag on the floor in order to get it to drain.It was stated that this was a concern when doing surgery as the provider were finding full bladders possible leading to one injury so far.While putting the bag on the floor was helping, it was absolutely against all infection control standards.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "static/ positive air pressure".It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.The product catalog number for this device was unknown.Therefore, bd was unable to determine the associated labeling to review.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the customer had multiple complaints about the surestep foley tray system that they had in stock.Stated that the main concern was foley bag not draining effectively and the nurses have to put the bag on the floor in order to get it to drain.It was stated that this was a concern when doing surgery as the provider were finding full bladders possible leading to one injury so far.While putting the bag on the floor was helping, it was absolutely against all infection control standards.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URINE COLLECTION
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14135657
Report Number1018233-2022-02621
Device Sequence Number6291583
Product Code KNX
UDI-Device Identifier00801741074073
UDI-Public(01)00801741074073
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/07/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberA902916
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer05/10/2022
07/07/2022
Initial Report FDA Received Date04/18/2022
Supplement Report FDA Received Date06/01/2022
07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-