ZIMMER BIOMET, INC. COMPR SRS MOD STEM - 6X75MM; SHOULDER, PROSTHESIS Back to Search Results
Catalog Number 211230
Medical Device Problem Codes Insufficient Device Problem Information (3190); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 09/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).It is unknown if product will be returning to zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.H3 other text : product not returned.
 
Event or Problem Description
It was reported patient is being considered for a revision due to unknown reasons.No revision has been reported to date.Attempts have been made and additional information on the reported event is unavailable.
 
Event or Problem Description
It was reported that the patient is being considered for a revision due to humeral loosening.Attempts have been made and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a3, b1, b3, b4, b5, b7, d2, d6, g1-2, g3, g7, h1, h2, h6, h10.Proposed annex g code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record was reviewed and no discrepancies were found.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: faint radiolucency along the humeral implant as noted which may be loose.Regarding alignment, there is slight humeral stem angulation as noted.Bone quality appears mildly osteopenic.The reported event is confirmed by mmi review.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPR SRS MOD STEM - 6X75MM
Common Device Name
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17388146
Report Number0001825034-2023-01715
Device Sequence Number6302382
Product Code JDC
UDI-Device Identifier00880304507791
UDI-Public(01)00880304507791(17)291203(10)243250
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number211230
Device Lot Number243250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2023
Supplement Date Received by Manufacturer08/01/2023
Initial Report FDA Received Date07/25/2023
Supplement Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age22 YR
Patient SexFemale
-
-