• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MICROSTREAM CO2 EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MICROSTREAM CO2 EXTENSION Back to Search Results
Model Number M3015A
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Unspecified Heart Problem (4454)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported that on (b)(6) 2022 at around 2 pm, there was a co2 equipment malfunction.A patient had a heart episode while connected to the device.The m3015a device was swapped out and sent for bench repair.Bench repair determined that the unit contains non-serviceable parts and the entire unit will require exchange.
 
Additional Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Additional Manufacturer Narrative
The catalog number, device name, fda product code and 510k have been corrected in this report.
 
Event or Problem Description
The customer reported that on (b)(6) 2022 at around 2 pm, there was a co2 equipment malfunction.A patient had a heart episode while connected to the device.The m3015a device was swapped out and sent for bench repair.Bench repair determined that the unit contained non-serviceable parts and the entire unit required exchange.A good faith effort (gfe) was made to determine whether the device caused or contributed to the patient heart episode.However, no further information was provided by the customer.The device had a confirmed malfunction.The investigation was unable to determine any relationship of the device malfunction to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSTREAM CO2 EXTENSION
Common Device Name
MICROSTREAM CO2 EXTENSION
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
jeanne ahearn
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15751134
Report Number9610816-2022-00559
Device Sequence Number6317499
Product Code DSI
UDI-Device Identifier00884838021556
UDI-Public00884838021556
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K993383
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM3015A
Device Catalogue NumberM3015A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date11/08/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexMale
-
-