• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 75447540
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age or date of birth: other- planned intervention.Product identifier is unknown, hence 510k# is not available.The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Radiographic image review stated that in the provided intraop x-rays, there is a fractured screw at s1.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient previously implanted with a screw in a normal spinal degenerative fixation at 4 levels.It was reported that the screw broke.The patient experienced pain.The event led to/ extended patient hospitalization for reintervention to remove the broken screw.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
A2: updated patient age a4: updated patient weight b2: updated outcomes attributed to adverse event b5: additional event details d2: updated fda product code d4: updated product identifiers d6a: updated explant date additional annex f code post device explantation g4: updated 510k# h6: updated eval.Code method as customer refused to return the product medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient previously implanted with a screw in a normal spinal degenerative fixation at 4 levels.It was reported that the screw broke.The patient experienced pain, but neither the patient nor the doctors knew that the screw was broken.The event led to/ extended patient hospitalization for reintervention to remove the broken screw.The patient achieved solid fusion.The date of implant was (b)(6) 2021 and the date of explant was (b)(6) 2022.No further complications were reported/ anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key15672064
Report Number1030489-2022-00976
Device Sequence Number6348047
Product Code KWP
UDI-Device Identifier00885074134208
UDI-Public00885074134208
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75447540
Device Catalogue Number75447540
Device Lot NumberH5549759
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date11/16/2022
Date Device Manufactured07/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age56 YR
Patient SexFemale
Patient Weight75 KG
-
-