• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCURAY INCORPORATED ACCURAY PRECISION TREATMENT PLANNING SYSTEM; SYSTEM, PLANNING, RADIATION THERAPY TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCURAY INCORPORATED ACCURAY PRECISION TREATMENT PLANNING SYSTEM; SYSTEM, PLANNING, RADIATION THERAPY TREATMENT Back to Search Results
Model Number TOMOTHERAPY TREATMENT DELIVERY SYSTEM WITH IDMS
Medical Device Problem Code Loss of Data (2903)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event may occur at healthcare facilities using at least 2 tomotherapy treatment delivery systems, when transferring a patient from one tomotherapy system machine to another under specific circumstances (patient transferred after an error occurred during fraction loading and user shuts down the machine before the treatment is initiated).There has been no patient death or serious injury.The root cause is an anomaly in software design: the unload fraction processed by the software did not require that unload should only be permitted from the treatment machine where the fraction was originally downloaded.The following action is being implemented: software patch installation to correct the anomaly.This event was initially considered to be part of mfr.Report# 3003873069-2021-00005.After additional assessment this is a separate event and therefore being reported separately.
 
Event or Problem Description
A fraction was delivered and completed normally without errors, but the system did not mark the fraction as completed.
 
Additional Manufacturer Narrative
During the investigation, it was initially determined that the root cause was the tomotherapy treatment delivery system with idms, however further investigation determined that the issue is with the precision treatment delivery system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCURAY PRECISION TREATMENT PLANNING SYSTEM
Common Device Name
SYSTEM, PLANNING, RADIATION THERAPY TREATMENT
Manufacturer (Section D)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer (Section G)
ACCURAY INCORPORATED
1209 deming way
madison WI 53717
Manufacturer Contact
daniel biank
1240 deming way
madison, WI 53717
6088242971
MDR Report Key13449861
Report Number3003873069-2022-00004
Device Sequence Number6552063
Product Code MUJ
UDI-Device IdentifierM65810682830
UDI-Public+M65810682830/16D20200219O
Combination Product (Y/N)N
Initial Reporter CountryGR
PMA/510(K) Number
K212794
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTOMOTHERAPY TREATMENT DELIVERY SYSTEM WITH IDMS
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2021
Supplement Date Received by Manufacturer10/13/2021
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-