ALLERGAN (COSTA RICA) STYLE 110 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 110-240
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Failure of Implant (1924); Seroma (2069)
Date of Event 06/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported left side rupture and textured implant.Device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.
 
Event or Problem Description
Healthcare professional additionally reported "fluid collection for prosthesis subcapsular area." device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed two openings assessed as an unidentified (tear) opening (shell thickness whiting to spec).Anxiety-product/procedure: unable to observe as it is a medical event not related to the device.Seroma-late: unable to observe as it is a medical event not related to the device.Additional observations: crease flat observed on the device.No further actions are required as the device was implanted.
 
Event or Problem Description
Patient reported left side rupture and textured implant.Healthcare professional additionally reported "fluid collection for prosthesis subcapsular area." device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 110 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15047575
Report Number9617229-2022-12386
Device Sequence Number6599180
Product Code FTR
UDI-Device Identifier10888628002203
UDI-Public10888628002203
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/29/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/29/2020
Device Model Number110-240
Device Catalogue Number110-240
Device Lot Number2734870
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer07/21/2022
11/04/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date08/17/2022
11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient RaceAsian
-
-