• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS INTEGRA GLUCOSE HK GEN. 3; HEXOKINASE, GLUCOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS INTEGRA GLUCOSE HK GEN. 3; HEXOKINASE, GLUCOSE Back to Search Results
Model Number C502
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The magnesium reagent lot number was 671588.The expiration date was not provided.The glucose reagent lot number was 698268.The expiration date was not provided.The customer performed the weekly maintenance and reran qc.Qc was acceptable.No further issues were reported afterward.
 
Event or Problem Description
We received an allegation about discrepant results for 1 patient's serum sample tested with a magnesium assay and with a glucose assay on a cobas 8000 cobas c502 module.Magnesium: initial result: 0.95 mg/dl rerun result: 1.6 mg/dl.Glucose: initial result : 37.1 mg/dl rerun result: 83.0 mg/dl.The repeat results were deemed to be correct.
 
Additional Manufacturer Narrative
Sections d, d1, d2, d2b, d4, g1 and g4 were updated.The investigation did not identify a product problem.The cause of the event could not be determined.The issue was resolved by performing the weekly maintenance (cleaning cell covers, rinse stations, sample and reagent probes, nozzles, and running the weekly pipe including washing reaction parts).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS INTEGRA GLUCOSE HK GEN. 3
Common Device Name
HEXOKINASE, GLUCOSE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key17175367
Report Number1823260-2023-02039
Device Sequence Number6617479
Product Code CFR
UDI-Device Identifier04015630920297
UDI-Public04015630920297
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K061048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberC502
Device Catalogue Number04404483190
Device Lot Number69826801
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2023
Supplement Date Received by Manufacturer06/30/2023
Initial Report FDA Received Date06/21/2023
Supplement Report FDA Received Date07/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-