• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227)
Date of Event 05/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight and ethnicity: information unknown/not provided.Asku.Device evaluation: product testing could not be performed as the product was not returned.The reported event could not be verified, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Conclusion: based on the investigation, no product deficiency was identified.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
It was reported that the jnj intraocular lens was explanted from the patient¿s left eye due to them experiencing halos and blurry vision.Their pre-operative visual acuity was 20/60 and post-operative 20/30.The incision had to be enlarged.There were no sutures used, no vitrectomy required and no delay in procedure.Reportedly, the patient is doing well after the explant.No further information was provided.
 
Additional Manufacturer Narrative
This follow-up report is to correct the initial submission.In the initial report the following fields were completed based on the wrong serial number provided by the customer.They've now clarified the correct serial number is (b)(6).The following fields are now updated accordingly.Section d4 additional device information model number: dfr00v, catalog number: dfr00vu190, serial number: (b)(6), expiration date: sep 27, 2024, unique identifier (udi) number: (b)(4), device manufacture date: sep 27, 2021.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: jan 12, 2023.Section h3: evaluated by manufacturer: yes.Device evaluation: the suspect iol was received cut with viscoelastic residue on the surface.The lens was cleaned and visual inspection found scratches on the lens surface.No other issues were identified.A product quality deficiency could not be determined.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15974338
Report Number3012236936-2022-02946
Device Sequence Number6960284
Product Code MFK
UDI-Device Identifier05050474750654
UDI-Public(01)05050474750654(17)240927
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/27/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer12/23/2022
01/27/2023
Initial Report FDA Received Date12/13/2022
Supplement Report FDA Received Date01/18/2023
02/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-