• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN 100 W/GRIPTION 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN 100 W/GRIPTION 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121731052
Medical Device Problem Codes Off-Label Use (1494); Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907)
Date of Event 10/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2012, the patient had a left hip arthroplasty to address osteoarthritis.Depuy components were implanted during this procedure.No mention of manufacturer of cement.On (b)(6) 2021, the patient had a revision left total hip to address mechanical complication of left hip prosthesis.The indications for surgery included, symptoms consistent with metallosis and excessive wear.During the procedure, the surgeon observed gross black tissue staining and a pseudotumor.The surgeon noted that the polyethylene shell completely broken free of the liner and the head was articulating with the cup.There was a large amount of titanium wear debris in the shell.The depuy acetabular cup was retained and the surgeon elected to use a bur to roughen the undersurface of the cup.The surgeon then roughened up the back of the new cup and cemented the competitor cup into the retained depuy cup.The stem, head, and acetabular liner were removed.Doi:(b)(6) 2012 dor: (b)(6) 2021 (left hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : on (b)(6) 2021, the patient had a revision left total hip to address mechanical complication of left hip prosthesis.The indications for surgery included, symptoms consistent with metallosis and excessive wear.During the procedure, the surgeon observed gross black tissue staining and a pseudotumor.The surgeon noted that the polyethylene shell completely broken free of the liner and the head was articulating with the cup.There was a large amount of titanium wear debris in the shell.The depuy acetabular cup was retained and the surgeon elected to use a bur to roughen the undersurface of the cup.The surgeon then roughened up the back of the new cup and cemented the competitor cup into the retained depuy cup.The stem, head, and acetabular liner were removed.Competitor components were implanted during this procedure.Doi:(b)(6) 2012 dor: (b)(6) 2021 (left hip).The product was not returned to depuy synthes, however radiological images were provided for review.The radiological images were reviewed, however they do not represent the reported complaint condition.The images show the competitors components that were implanted during the revision.Due to the radiopacity of the radiological images it is not possible to identify if there are two cups implanted on the left side of the hip.As per the acetabular cup's instruction for use (ifu-0902-00-701 rev.T), implants and trials components from different manufacturers or implant systems should never be used together.Additionally, single use devices manufactured by depuy have not been designed to undergo or withstand any form of alteration, such as disassembly, cleaning or re-sterilization, after a single patient use.Do not alter or modify implants in any way.No additional information was provided to confirm if the acetabular cup was altered to be assemble with a competitors cup.Therefore, no conclusion can be drawn and no further investigation is required since the implant was not returned for evaluation.A manufacturing record evaluation was performed for the finished device [121731052 / 120770b] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified during manufacturing.Based on the available information, the reported off label allegation was confirmed since it was reported that the inner surface of the pinn 100 w/gription 52mm was altered.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device [121731052 / 120770b] number, and no non-conformances / manufacturing irregularities related to the malfunction were identified during manufacturing.
 
Event or Problem Description
Medical records were received: on (b)(6) 2012, the patient had a left hip arthroplasty to address osteoarthritis.Depuy components were implanted during this procedure.No mention of manufacturer of cement.On (b)(6) 2021, the patient had a revision left total hip to address mechanical complication of left hip prosthesis.The indications for surgery included, symptoms consistent with metallosis and excessive wear.During the procedure, the surgeon observed gross black tissue staining and a pseudotumor.The surgeon noted that the polyethylene shell completely broken free of the liner and the head was articulating with the cup.There was a large amount of titanium wear debris in the shell.The depuy acetabular cup was retained and the surgeon elected to use a bur to roughen the undersurface of the cup.The surgeon then roughened up the back of the new cup and cemented the competitor cup into the retained depuy cup.The stem, head, and acetabular liner were removed.Competitor components were implanted during this procedure.On (b)(6) 2021 the medical records note the patient had multiple surgeries: 9 on the right hip, 3 on the left hip (no specific info).Patient reports having constant pain in the bilateral hips, lumbar spine, and cervical spine.The patient¿s impairments include pain, pelvic alignment issues and decreased function.On (b)(6) 2022, a rehab outpatient communication reports that the patient stated that they had metal poisoning.Doi: on (b)(6) 2012 dor: on (b)(6) 2021 (left hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN 100 W/GRIPTION 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17798423
Report Number1818910-2023-19307
Device Sequence Number7028745
Product Code KWA
UDI-Device Identifier10603295010203
UDI-Public10603295010203
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K090998
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/22/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/01/2021
Device Catalogue Number121731052
Device Lot Number120770B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Supplement Date Received by Manufacturer09/25/2023
12/15/2023
01/22/2024
03/06/2024
03/11/2024
03/26/2024
Initial Report FDA Received Date09/22/2023
Supplement Report FDA Received Date09/26/2023
12/21/2023
01/26/2024
03/07/2024
03/15/2024
04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/12/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 36IDX52OD; CORAIL AMT COLLAR SIZE 12; DELTA CER HEAD 12/14 36MM +1.5; UNK HIP ACETABULAR LINER POLY; UNK HIP FEMORAL HEAD METAL; UNK HIP FEMORAL STEM CORAIL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-