• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS C 503 ANALYTICAL UNIT; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS C 503 ANALYTICAL UNIT; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Catalog Number 08463662001
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The c503 analyzer serial number was (b)(6).Calibration and qc were acceptable.An "abnormal aspiration" message was observed on the alarm trace data around the time of the event.The hardware performance check results showed issues with cell rinse functionality.The field service engineer (fse) cleaned the probes as there were some particles on the sodium hydroxide (naoh) probe.The suction probe was dripping; the vacuum line was bleached.The investigation is ongoing.
 
Event or Problem Description
The initial reporter complained of a discrepant high result for 1 patient sample tested for triglycerides (trigl) on a cobas c 503 analytical unit.The initial result was 5.27 g/l.The sample was repeated twice with results of 1.00 g/l.The questionable result was not reported outside of the laboratory.The repeat results were believed to be correct.
 
Additional Manufacturer Narrative
Section d, device identification was updated.Relevant fields of sections d and g were updated.The trigl reagent lot number was 766131 with an expiration date of (b)(6)2024.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS C 503 ANALYTICAL UNIT
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250-0457
MDR Report Key18641006
Report Number1823260-2024-00321
Device Sequence Number7342928
Product Code JJE
UDI-Device Identifier07613336158784
UDI-Public07613336158784
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K191899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number08463662001
Device Lot Number766131
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/12/2024
Supplement Date Received by Manufacturer02/27/2024
Initial Report FDA Received Date02/05/2024
Supplement Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age20 YR
Patient SexFemale
-
-