• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS MG2; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number C501
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Na.
 
Event or Problem Description
The initial reporter complained of questionable results for multiple patients tested for mg2 on a cobas 6000 c (501) module.Discrepant results were provided for 1 patient sample.The initial result was 3.9 mg/dl.This result was reported outside of the laboratory where it was questioned.The repeat result was 2.2 mg/dl.The repeat result was believed to be correct.The mg2 reagent lot number was 596601.The expiration date was not provided.
 
Additional Manufacturer Narrative
The c501 module serial number was (b)(6).No pre-analytic issues were identified.Calibration was last performed on (b)(6) 2022 and was acceptable.Qc was acceptable.There was no indication for a performance issue of the reagent.No issues were identified during a review of the alarm trace data.The field service engineer (fse) found the reagent pack was contaminated and replaced it.The customer ran precision testing along with a patient correlation.All results were reproducible and precision results were within specification.The customer has had no further issues since the service visit.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MG2
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13929974
Report Number1823260-2022-00880
Device Sequence Number7566993
Product Code JGJ
UDI-Device Identifier04015630929818
UDI-Public04015630929818
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K954992
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC501
Device Catalogue Number06481647190
Device Lot Number596601
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-