• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO GIA; STAPLER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO GIA; STAPLER, SURGICAL Back to Search Results
Model Number GIA8048S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Purulent Discharge (1812); Fever (1858); Fistula (1862); Electrolyte Imbalance (2196); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/02/2016
Type of Reportable Event Serious Injury
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of division of colon approximation of rectal stump.It was reported that after the implant, the patient experienced air-containing collection, abscess, labs abnormal for creatinine, wbc, purulent fluid, fistula, fever, electrolyte imbalance, anastomosis leak.Post-operative patient treatment included ct scan, hospitalization, interventional radiology procedure, biopsy/aspiration of abdomen drainage, antibiotics, potassium supplementation.
 
Additional Manufacturer Narrative
Additional info: b2, b3, b5, b6, b7, d1, d2, d3, d4 (model #, catalog #, expiration date, lot #, unique identifier (udi) #), g4 (pma / 510(k) #), h4, & h6 (patient codes, device codes, ime e2402: air-containing collection, labs abnormal for creatinine; wbc).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an unknown problem.It was reported that after the implant, the patient experienced an unspecified adverse outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce,pr CT 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce,pr CT 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19864724
Report Number1219930-2024-03441
Device Sequence Number8074843
Product Code GAG
UDI-Device Identifier10884523007282
UDI-Public10884523007282
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K221013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model NumberGIA8048S
Device Catalogue NumberGIA8048S
Device Lot NumberP5D0775KX
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/30/2024
Supplement Report FDA Received Date08/05/2024
Date Device Manufactured04/29/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient SexUnknown
-
-