MAUDE Adverse Event Report: ABBOTT DIABETES CARE LTD LIBRE 3 APP FREESTYLE; CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE LTD LIBRE 3 APP FREESTYLE; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71926-01
Medical Device Problem Codes Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591); Application Program Problem (2880)
Health Effect - Clinical Code Convulsion/Seizure (4406)
Date of Event 02/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer reported a replace sensor error message on app screen.The reported issue was investigated and attempted to be replicated.Per the abbott diabetes care compatibility guide for the freestyle libre 2 app art41556-202, the reported configuration of ios 17.2.1 is not compatible with the freestyle libre 2 app.The latest revision of the compatibility guide is available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.There were no issues identified with the freestyle libre 2 app during replication that would have led to the reported issue.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A "replace sensor" error message was reported with the abbott diabetes care (adc) device using the freestyle libre 2 application on the customer¿s iphone 13 with ios operating system version 17.2.1 and the customer was unable to obtain readings.The customer experienced seizure, and no third-party treatment was reported.No further details were provided.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
An extended investigation has been performed for the reported complaint.The customer reported an error message on app screen.The reported issue was investigated and attempted to replicate.The reported configuration was not compatible with the freestyle libre 3 app.The latest revision of the compatibility guide was available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.This serves as a correction report.Section d1 (brand name) and h10 (addtl mfg narrative) were incorrectly documented in the initial report.The correction has been made here.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A "replace sensor" error message was reported with the abbott diabetes care (adc) device in use with iphone 13, os 17.2.1, app version 3.5.0.8863 and the customer was unable to obtain readings.The customer experienced seizure, and no third-party treatment was reported.No further details were provided.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned and a valid serial number has not been provided.Clinical data was reviewed and confirmed that libre sensors continue to be safe, effective, and perform as intended in the field.A tripped trend review was completed for the reported complaint and fs libre sensors, and the investigation concluded that the tripped trend was not correlated with a product nonconformance.An extended investigation has been performed for the reported complaint.The customer reported an error message on app screen.The reported issue was investigated and attempted to replicate.The reported configuration was not compatible with the freestyle libre 3 app.The latest revision of the compatibility guide was available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.H10 was incorrectly documented in the previous mdr submission.H11 has been updated to reflect correct investigation activities.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
A "replace sensor" error message was reported with the abbott diabetes care (adc) device in use with iphone 13, os 17.2.1, app version 3.5.0.8863 and the customer was unable to obtain readings.The customer experienced seizure, and no third-party treatment was reported.No further details were provided.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBRE 3 APP FREESTYLE
Common Device Name
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney CA 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18864460
Report Number2954323-2024-08057
Device Sequence Number8239991
Product Code QBJ
UDI-Device Identifier00357599810009
UDI-Public00357599810009
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K223435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/04/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71926-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2024
Supplement Date Received by Manufacturer03/11/2024
05/15/2024
Initial Report FDA Received Date03/08/2024
Supplement Report FDA Received Date03/28/2024
06/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-