• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE INDUSTRIES, LP; 3ML SYRINGE IN PACK EYE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE INDUSTRIES, LP; 3ML SYRINGE IN PACK EYE Back to Search Results
Model Number DYNJ60113C
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that a syringe component was noted to have black staining within the barrel.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation.The sample was filled with liquid and inside there appeared to be a dark greyest stain or material that appeared to be embedded with the material of the syringe.In an abundance of caution, and in response to an fda 483 issued for cfn 1417592 on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that a syringe component was noted to have black staining within the barrel.
 
Additional Manufacturer Narrative
Update made to d2 product information.
 
Event or Problem Description
It was reported that a syringe component was noted to have black staining within the barrel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
3ML SYRINGE IN PACK EYE
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18881117
Report Number1423395-2024-00317
Device Sequence Number9255558
Product Code FMF
UDI-Device Identifier40195327177479
UDI-Public40195327177479
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/28/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDYNJ60113C
Device Lot Number23IMH496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date05/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-