SMITH & NEPHEW, INC. GII ARTICULAR INSERTER/EXTRACT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
71440194 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that (2) 2.5mm hex holding sleeve 2cm have broken captures.It was unknown if this occurred during surgery or not; therefore, patient involvement has not been confirmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections b3, d1, d2 and d4 were corrected due to new information received.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tka (2) gii articular inserter/extract, presented break captures.The surgery was completed without delay with a backup device.No patient injuries or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failure for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instruments that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|