WILLIAM COOK EUROPE COOK CELECT NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Catalog Number IGTCFS-65-1-FEM-CELECT |
Device Problems
Structural Problem (2506); Device Tipped Over (2589); Insufficient Information (3190)
|
Patient Problems
Pain (1994); Thrombosis (2100); Perforation of Vessels (2135); Device Embedded In Tissue or Plaque (3165); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2014.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
|
|
Event Description
|
Patient allegedly received an implant in (b)(6)2014 via the right common femoral vein due to post deep vein thrombosis (dvt).The patient alleges tilt and vena cava perforation.The patient further alleges physical pain, limited activity, and dvt.(b)(6) 2018, per a report from computed tomography; ¿some of the filter struts have penetrated through the wall of the ivc into the pericaval/mesenteric fat.The filter is tilted to the right with its proximal cone lying on the wall of the ivc possibly embedded in it.¿ (b)(6), per a report from ct; ¿the infrarenal ivc filter is in place, slightly tilted to the right.The legs of the ivc filter appeared to traverse the wall of the ivc.There is otherwise no obvious fracture.Compared to the radiograph dated (b)(6)2018, this is unchanged in alignment.(b)(6) 2019, per a report from xray; ¿an ivc filter is again seen at l2 - 3 level.¿.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation the following allegations have been investigated: vena cava perforation, deep vein thrombus, tilt, pain, limited activity.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain and limited activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: b5, b6, d6b, h6.Investigation the following allegations have been investigated: embedment.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The additional information regarding embedment does not change the previous investigation results for vena cava perforation.(b)(4) devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Additional information has been provided.The patient alleges embedment and a successful percutaneous filter retrieval on (b)(6) 2020.11jun2020, per a report from retrieval report (successful); ¿we then delivered the sheath over the ivc filter to collapse the struts and successfully retracted the filter to within the sheath.We then removed the entire system as a unit while visualizing that the ivc filter remained contained within the sheath.¿.
|
|