Concomitant medical products: other relevant device(s) are: product id: 9731203, serial/lot #: (b)(4), udi#: (b)(4).A medtronic representative went to the site to perform a system check out and they replaced the emitter.The system functioned as intended.Fdm10, fdr120, fdc4307 are applicable to the system checkout.The electromagnetic field generator was returned for product analysis.The reported problem could not be duplicated.The field generator was connected to a test system for a burn-in test.The system remained in green status during all testing.Flexing the cable does not indicate any intermittent opens.No problem was found, the system was fully functional.Fdm10, fdr213, fdc67 are applicable to the product analysis.If information is provided in the future, a supplemental report will be issued.
|