| Model Number |
ETLW1628C124E |
| Medical Device Problem Codes |
Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other relevant device(s) are: product id: e sbf3214c103ee, serial/lot #: (b)(4), ubd: 21-feb-2019, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant iis bifurcate stent graft system was implanted in the endovascular treatment of a 50mm abdominal aortic aneurysm.It was reported that a ct performed just over two years later showed a type iii endoleak.As per the physician it is thought that the limb etlw1628c124e has separated from the main body stent graft.As per the physician the cause of the event can not be determined.No additional clinical sequalae were provided and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
B:5 additional information received: it was confirmed that the stent graft limbs (b)(6) and (b)(6) had separated.A secondary procedure was carried out five days post when the endoleak was observed on ct, and the stent graft limbs (b)(6) and (b)(6) were implanted as treatment.Follow up ct showed that the endoleak is resolved.There is now no complaint against pli-10 (v07573832) and pli-20 (v07333222) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
B:5 additional information received; the patient had a ct performed which showed an internal iliac aneurysm.It was said in the previous intervention procedure, the physician had coiled the left internal iliac at the time but the aneurysm continued to grow since.As a result an endurant limb was implanted on the same date of the ct to cover internal and extend into the external iliac artery.This resolved the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|