The reported event was confirmed, as manufacturing-related.The amber lubricath 75cc irrigation foley catheter with at9268 date code was inflated with 80cc of water and no leaks were noted.The balloon was then passively deflated.Water was found to remain in the balloon after deflation.This did not meet specifications.The device appeared to be used for treatment.As the balloon did not fully deflate there appeared to be a relationship with between the device and the reported event.A potential root cause could be due to "operator error", "insufficient drying", or "dulling cutting tool".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|