| Model Number |
AIRLIFE MISTY-NEB MEDICATION NEBULIZERS |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No samples were returned as of now.A second nebulizer was obtained and set up.Medication transferred and misted appropriately on the second nebulizer.The customer confirmed that no harm occurred to the patient.Any additional information received from the customer will be included in a follow-up report.
|
| |
|
Event or Problem Description
|
|
The customer reported that the airlife misty max 10 nebulizer would not mist.The customer confirmed that there was no patient harm associated with this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: review of device history record of the lot number revealed no anomalies, and a sample had been received by the manufacturer for evaluation.Unfortunately, that sample was discarded by mistake.However, technical team indicated that capa-000000567 has been initiated based on a similar problem, and the manufacturing date of the product reported on this complaint is covered by that capa.Therefore, technical team is confirming the reported failure.
|
| |
|